SATEC PM180 Operation Manual User Manual

Page 92

Advertising
background image

‎Chapter 8 Configuring Recorders

Configuring the IEEE 1159 Power Quality Recorder

90

PM180 Substation Automation Unit

used to detect voltage and frequency faults, their typical triggering thresholds and
event durations.
Impulsive transients

Impulsive transients are detected as impulses with a rise time less than 0.5 ms and
duration from 150 us to ½ cycle. The impulse magnitude is measured as the
overshoot voltage magnitude above the normal voltage wave shape. It is referenced
to the nominal peak voltage (1.414 Un).
Sags and swells

A voltage sag or swell is classified as one polyphase event regardless of the shape
and of the number of phases affected (as per IEC 61000-4-30). An event can begin
on one phase and end on another phase. The fault magnitude is recorded separately
for each phase involved. The event duration is measured from the instant at which
the voltage falls/rises below/above the start threshold on one of the phases to that at
which it becomes greater/lower than the end threshold on all affected phases
including a threshold hysteresis.
Voltage Interruptions

The voltage interruption is detected when the voltages on all phases fall below the
interruption threshold (as per IEC 61000-4-30).
Flicker

The flicker is evaluated according to IEC 61000-4-15. The 10-min Pst values are
used to follow and indicate the flicker faults.

Event ID IEEE 1159 category

Trigger parameter

Reference value Typical

thresholds, %

Event
duration

PQE11

Impulsive transients

Instantaneous overshoot voltage

Un peak voltage

20-200%

150 us

–10 ms

PQE211

Instantaneous sag

½ cycle RMS voltage

Un RMS

80-90%

< 30 cycles

PQE212

Instantaneous swell

½ cycle RMS voltage

Un RMS

110-120%

< 30 cycles

PQE221

Momentary interruption

½ cycle RMS voltage

Un RMS

0-10%

< 3 sec

PQE222

Momentary sag

½ cycle RMS voltage

Un RMS

80-90%

< 3 sec

PQE223

Momentary swell

½ cycle RMS voltage

Un RMS

110-120%

< 3 sec

PQE231

Temporary interruption

½ cycle RMS voltage

Un RMS

0-10%

< 1 min

PQE232

Temporary sag

½ cycle RMS voltage

Un RMS

80-90%

< 1 min

PQE233

Temporary swell

½ cycle RMS voltage

Un RMS

110-120%

< 1 min

PQE31

Sustained interruption

½ cycle RMS voltage

Un RMS

0-10%

> 1 min

PQE32

Undervoltage

½ cycle RMS voltage

Un RMS

80-90%

> 1 min

PQE33

Overvoltage

½ cycle RMS voltage

Un RMS

110-120%

> 1 min

PQE4

Voltage unbalance

3-sec negative sequence unbalance No

1-5%

Steady state

PQE52

Harmonics THD

3-sec harmonic THD

No

5-20%

Steady state

PQE53

Interharmonics THD

3-sec interharmonic THD

No

2-8%

Steady state

PQE6

Voltage fluctuations (flicker) 10-min Pst

No

1-5

Steady state

PQE7

Frequency variations

3-sec frequency

Nominal frequency

1-6%

Steady state

Un

– nominal device voltage

PQ Recorder Setup

The IEEE 1159 PQ recorder setup allows you to adjust thresholds and hysteresis for
PQ triggers, to define the waveform and data log options for PQ events, and to
enable or disable the PQ recorder in your device.
To configure the PQ recorder:

1. Select Memory/Log from the Meter Setup menu, and then click on

the PQ Recorder tab.

2. If you want to change the default settings, adjust thresholds and

hysteresis for PQ triggers

3. Select the waveform and data logging options for PQ events
4. Download your setup to the device.

Advertising