Proportional counts to engineering units, Engineering units to proportional counts – Rockwell Automation 1746-NR8 SLC 500 RTD/Resistance Input Module User Manual User Manual

Page 60

Advertising
background image

Publication 1746-UM003A-EN-P

4-12 Channel Configuration, Data, and Status

Proportional Counts to Engineering Units

Equation:

Assume that input type is a potentiometer (1000

, range = 0 to 1000

),

proportional counts display type. Channel data = 21567.

Want to calculate ohms equivalent.

From Channel Data Word Format, S

LOW

= 0

and S

HIGH

= 1000

.

Solution:

Engineering Units to Proportional Counts

Equation:

Assume that input type is a potentiometer (3000

, range = 0 to 3000

),

proportional counts display type. Desired channel resistance value = 1809

.

Want to calculate Proportional Counts equivalent.

From Channel Data Word Format, S

LOW

= 0

and S

HIGH

= 3000

.

Solution:

The following table shows the temperature ranges of several 1746-NR8 RTDs.
The table applies to both 0.25 and 1.0 mA excitation currents. The
temperature ranges of the remaining RTDs vary with excitation current, for
example, 1000

Platinum 385, 1000

Platinum 3916

,

and 10

Copper 426.

Engr Units Equivalent

SLOW

SHIGH - SLOW

(

)

Proportional Counts value displayed + 32768

(

)

65536

-------------------------------------------------------------------------------------------------------------------------

×

+

=

Engr Units Equivalent

0

1000 ohms - 0 ohms

[

]

21567 32768

+

(

)

65536

--------------------------------------------

×

826 ohms

=

+

=

Proportional Counts Equivalent

65536

Engr Units desired - SLOW

(

)

SHIGH - SLOW

(

)

--------------------------------------------------------------------------

×

- 32768

=

Prop. Counts = {65536 x [1809 ohms - 0 ohms]} -32768 = 6750