Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 102

Advertising
background image

5.12 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

92

■■■■■■■■

881 Compact IC pro – Anion – MCS

Replacing parts of the suppressor

Replace the parts of the suppressor as follows:

1 Disconnecting the suppressor from the IC system

Switch off the instrument.

Disconnect all capillaries of the suppressor from the IC system.

2 Dismantling the suppressor

Unscrew union nut (42-1) from the housing (42-4).

Pull the connecting piece (42-2) and the rotor (42-3) out of the
housing.
The connecting piece and the rotor normally stick to one another
– if this is not the case: Take a sharp object, insert it into the slot
of the housing (42-5) and pull out the rotor in this way.

Detach the connecting piece from the rotor.

3 Cleaning the new rotor

Clean the surface of the new rotor (42-3) with ethanol using a
lint-free cloth.

4 Inserting the new rotor

Caution

An incorrectly inserted rotor can be destroyed during start-up.

Insert the new rotor (42-3) into the housing (42-4) in such a way
that the tubing connections on the rear of rotor fit into the corre-
sponding recesses inside the housing and one of the three holes
of the rotor is visible from below in the slot of the housing (42-5).

If the rotor is correctly inserted, its sealing surface will be approx.
4 mm within the housing. If this is not the case, the rotor must be
moved into the right position from below using a sharp object.

5 Cleaning the new connecting piece

Clean the surface of the new connecting piece (42-2) with etha-
nol using a lint-free cloth.

Advertising