5.11 Peristaltic pump

■■■■■■■■■■■■■■■■■■■■■■

86

■■■■■■■■

881 Compact IC pro – Anion – MCS

Order
number

Name

Material

Inner
diameter

Use

6.1826.360

Pump tubing LFL
(white/white), 3-stop-
per

PVC (Tygon)

1.02 mm

For sample transfer.

6.1826.380

Pump tubing LFL
(gray/gray), 3-stopper

PVC (Tygon)

1.25 mm

For inline sample dilution.

6.1826.390

Pump tubing LFL (yel-
low/yellow), 3-stop-
per

PVC (Tygon)

1.37 mm

For the sample solution in
inline ultrafiltration.

5.11.2.2

Pump tubing connection with filter

The 6.2821.130 filters (41-2) should be changed every 3 months, more
frequently at higher backpressure.

1

2

3

Figure 41

Pump tubing connection – Changing the filter

1

Tubing olive

2

Filter 6.2821.130
Packaging contains 10 items.

3

Filter housing

Replacing the filter

1 Unscrewing filter screw

Screw the filter screw (41-3) out of the tubing olive (41-1) with
the aid of two 6.2621.000 adjustable wrenches.

2 Replacing the filter

Remove the old filter (41-2) with tweezers.

Place the new filter (41-2) flat in the tubing olive (41-1) with
tweezers.

3 Mounting filter screw

Screw the filter screw (41-3) back into the tubing olive (41-1) and
tighten by hand. Then additionally tighten with two 6.2621.000
adjustable wrenches.