Figure 28, Suppressor – connection capillaries – Metrohm 881 Compact IC pro – Anion – MCS User Manual
Page 62
3.17 Metrohm Suppressor Module (MSM)
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
881 Compact IC pro – Anion – MCS
1
3
2
3
4
5
6
7
8
2
1
Figure 28
Suppressor – connection capillaries
1
Union nut
2
Connecting piece (6.2832.010)
3
Eluent inlet capillary
Labeled with in.
4
Eluent outlet capillary
Labeled with out.
5
Rinsing solution inlet capillary
Labeled with rinsing solution.
6
Rinsing solution outlet capillary
Labeled with waste rins..
7
Regeneration solution outlet capillary
Labeled with waste reg..
8
Regeneration solution inlet capillary
Labeled with regenerant.
The rinsing solution and regeneration solution are pumped with a peristal-
tic pump (see Chapter 3.16, page 45).
Caution
To protect the suppressor against foreign particles or bacterial growth,
a pump tubing connection with filter (6.2744.180) (26-3) must be
mounted between the peristaltic pump and the inlet capillaries of the
suppressor.