3.17 Metrohm Suppressor Module (MSM)

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

881 Compact IC pro – Anion – MCS

1

3

2

3

4

5

6

7

8

2

1

Figure 28

Suppressor – connection capillaries

1

Union nut

2

Connecting piece (6.2832.010)

3

Eluent inlet capillary
Labeled with in.

4

Eluent outlet capillary
Labeled with out.

5

Rinsing solution inlet capillary
Labeled with rinsing solution.

6

Rinsing solution outlet capillary
Labeled with waste rins..

7

Regeneration solution outlet capillary
Labeled with waste reg..

8

Regeneration solution inlet capillary
Labeled with regenerant.

The rinsing solution and regeneration solution are pumped with a peristal-
tic pump (see Chapter 3.16, page 45).

Caution

To protect the suppressor against foreign particles or bacterial growth,
a pump tubing connection with filter (6.2744.180) (26-3) must be
mounted between the peristaltic pump and the inlet capillaries of the
suppressor.