6.1 Problems and their solutions

■■■■■■■■■■■■■■■■■■■■■■

98

■■■■■■■■

881 Compact IC pro – Anion – MCS

Problem

Cause

Remedy

Sample – blockage in the
sample path.

Check the sample path.

Sample – sample loop not
(completely) filled.

Prolong the sample transfer time.

Sample – gas bubbles in
the sample.

Use sample degasser (see Chapter 3.13, page
37)
.

MCS – not connected.

Connect the MCS.

Peristaltic pump –
insufficient or no
delivery rate

Peristaltic pump – contact
pressure too weak.

Correctly set contact pressure (see "Set flow
rate", page 50)
.

Peristaltic pump – filter
blocked.

Replace the filter (see Chapter 5.11.2.2, page
86)
.

Peristaltic pump – pump
tubing defective.

Replace pump tubing (see Chapter 5.11.2.1,
page 85)
.

Very noisy baseline

High pressure pump – con-
taminated pump valves.

Clean pump valves (see Chapter 5.5.2, page
70)
.

Eluent – Leak in the eluent
path.

Check the eluent path

Eluent – Blockage in the
eluent path.

Check the eluent path

High pressure pump –
defective piston seals.

Replace (see Chapter 5.5.2, page 70) piston
seals .

MCS –
CO

2

adsorption cartridge

exhausted.

Replace CO

2

adsorption cartridge (see Chapter

5.13.1, page 93).

Pulsation absorber not
connected. or defective.

Connect pulsation absorber (see Chapter 3.12,
page 36)
. or replace it.

MCS – vacuum pump
defective.

Contact the Metrohm Service.

Background conduc-
tivity too high

Suppressor – not connec-
ted.

Connect the suppressor (see Chapter 3.17.1,
page 51)
.

MCS – not connected.

Connect the MCS.

Incorrect eluent.

Change eluent (see Chapter 5.4.2.3, page 69).