3.6 Rear of the instrument

■■■■■■■■■■■■■■■■■■■■■■

20

■■■■■■■■

881 Compact IC pro – Anion – MCS

Colored sleeves for PEEK capillaries
The enclosed set of varicolored sleeves for PEEK capillaries (6.2251.000)
serves to easily differentiate the various flows of liquid in the system
through color coding. Each capillary leading a given liquid (e. g. eluent)
can be highlighted with sleeves of the same color.

To highlight a capillary, proceed as follows:

1 Slide a sleeve of a selected color over a capillary an move it to an

easily visible position.

If the capillary heats up, the sleeve shrinks and adapts to the form of
the capillary.

3.6

Rear of the instrument

3.6.1

Transport locking screws

To avoid damage to the high pressure pump and vacuum pump during
transport, the pumps are secured with transport locking screws .

Remove these transport locking screws before the initial start-up.

Removing transport locking screws

1 Remove all of the transport locking screws with the 6.2621.030 4

mm hexagon key and keep them in a safe place.

Warning

In order to avoid damage to the pump, the transport locking screws
must be remounted each time the instrument undergoes major trans-
port.

3.6.2

Leak sensor

The leak sensor detects escaping liquid which collects in the base tray of
the instrument.

To activate the leak sensor, the leak sensor connector plug (5-2) must be
connected, the instrument switched on and the leak sensor switched to
active in MagIC Net.