Figure 29, Mcs – connection – Metrohm 881 Compact IC pro – Anion – MCS User Manual
Page 65
■■■■■■■■■■■■■■■■■■■■■■
3 Installation
881 Compact IC pro – Anion – MCS
■■■■■■■■
55
1
2
3
4
4
5
5
6
7
Figure 29
MCS – connection
1
MCS input
Connection to the MSM.
2
MCS output
Connection to the detector.
3
Aspirating capillary
For aspirating air low in CO
2
(as a result of
CO
2
adsorption cartridge (30-
)).
4
PEEK pressure screw, long
(6.2744.090)
5
Capillary connection
6
Pressure screw, short (6.2744.070)
Mounted on the air aspirating capillary.
7
Luer coupling (6.2744.120)
Mounted on the air aspirating capillary with
pressure screw 6.2744.070.
Connecting the MCS
1 Connecting the MSM
Connect the eluent outlet capillary (labeled with out) to the MCS
inlet (29-1) using a long PEEK pressure screw (6.2744.090) (29-4).
2 Connection to the detector
Connect the detector inlet capillary to the MCS output (29-2) using a
long PEEK pressure screw (6.2744.090) (29-4).
Caution
If the MCS is not used, the input and output must be sealed with
6.2744.220 stoppers.