6 troubleshooting, 1 problems and their solutions, Problems and their solutions – Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 107

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Troubleshooting

881 Compact IC pro – Anion – MCS

■■■■■■■■

97

6 Troubleshooting

6.1

Problems and their solutions

Problem

Cause

Remedy

Marked drop in
pressure

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.5, page 17).

Marked rise in pres-
sure

Inline filter (6.2821.120)
blocked.

Replace the filter (6.2821.130) (see Chapter
5.6, page 80)
.

Suppressor – blocked.

Regenerate the suppressor (see Chapter
5.12.3.2, page 88)
.

Note: 6.2821.180 pump tubing connection
with filter must be used.

Guard column – blocked.

Replace guard column (see Chapter 3.20,
page 59)
.

Separation column –
blocked.

Regenerate separation column (see Chap-
ter 5.14.4, page 95)
.

Replace separation column (see "Connect-
ing and rinsing the separation column",
page 61)
.

Note: Samples should always be microfiltered
(see Chapter 5.8, page 82).

Injection valve – valve
blocked.

Have the valve cleaned (by Metrohm service
technicians).

Drift of the baseline

Thermal equilibrium not yet
attained.

Condition instrument with the column heater
(see Chapter 3.15, page 41) switched on (see
Chapter 4.2, page 64)
.

Leak in the system.

Check all capillary connections and seal leaks,
if necessary (see Chapter 3.5, page 17).

Eluent – Evaporation of the
organic solvent in the elu-
ent.

Check the eluent bottle attachment (see Figure
10, page 27)
.

Peak areas lower
than expected

Sample – leak in the sam-
ple path.

Check the sample path.

Advertising