5 Operation and maintenance

881 Compact IC pro – Anion – MCS

■■■■■■■■

81

Changing the filter

The flow must be stopped before changing the filter.

1 Removing the inline filter

Unscrew the pressure screws (40-1) from the inline filter.

2 Unscrewing the filter screw

Screw the filter screw (40-4) out of the filter housing (40-2) with
the aid of two adjustable wrenches (6.2621.000).

3 Inserting the filter

Remove the old filter (40-3) using tweezers.

Place the new filter (40-3) flat in the filter housing (40-2) using
tweezers.

4 Mounting filter screw

Screw the filter screw (40-4) back into the filter housing (40-2)
and tighten by hand. Then additionally tighten slightly with two
adjustable wrenches (6.2621.000).

5 Remounting the inline filter

Screw pressure screws (40-1) back onto the inline filter.

6 Rinsing the inline filter

Dismantle the guard column (if present) and the separation col-
umn and replace with a coupling (6.2744.040).

Rinse the instrument with eluent.