3 Installation

881 Compact IC pro – Anion – MCS

■■■■■■■■

25

3.8

Eluent

3.8.1

Connecting eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing (8-1).

The eluent aspiration tubing is connected to the eluent degasser (see
Chapter 3.9, page 29)
. The tubing must be threaded through a suitable
capillary feed-through of the instrument before the other end can be
equipped.

You will require the parts from the following accessories for equipping the
eluent aspiration tubing:

6.1602.160 Eluent bottle attachment GL 45

6.2744.210 tubing adapter for aspiration filter

6.2821.090 aspiration filter

To equip the eluent aspiration tubing proceed as follows:

Assembling eluent aspiration tubing

1 Guide the free end of the eluent aspiration tubing (8-1) out of the

instrument through a suitable capillary feed-through.

2 Installing eluent bottle attachment 6.1602.160

Slide tubing nipple (8-2) and O-ring (8-3) onto the eluent aspira-
tion tubing (8-1).

Push eluent aspiration tubing (8-1) through the bottle attachment
(8-4) and screw tight.