5 Operation and maintenance

881 Compact IC pro – Anion – MCS

■■■■■■■■

89

Regenerating the suppressor

1 Disconnecting the suppressor from the IC system

Disconnect the capillaries of the suppressor labeled with regener-
ant
and rinsing solution from the IC system.

2 Connecting the suppressor to the high pressure pump

Connect the inlet capillary for the regeneration solution (labeled
with regenerant) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040)

3 Regenerating the suppressor

Regenerate the first suppressor unit for about 15 minutes.

In the software, use the Step command to switch to the second
suppressor unit and regenerate this for about 15 minutes.

In the software, use the Step command to switch to the third
suppressor unit and regenerate this for about 15 minutes.

4 Rinsing the suppressor

After regeneration, the three suppressor units must be rinsed with
degassed ultra pure water for about 15 minutes.

Remove the inlet capillary for the regeneration solution (labeled
with regenerant) from the outlet of the high pressure pump.

Connect the inlet capillary for the rinsing solution (labeled with
rinsing solution) to the outlet of the high pressure pump with the
aid of a coupling (6.2744.040).

Rinse the first suppression unit with degassed ultra pure water for
about 15 minutes.

In the software, use the Step command to switch to the second
suppressor unit and rinse this for about 15 minutes.

In the software, use the Step command to switch to the third
suppressor unit and rinse this for about 15 minutes.

5 Connect the suppressor to the IC system

Reconnect the capillaries labeled regenerant and rinsing solution
to the IC system.