Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 67

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

881 Compact IC pro – Anion – MCS

■■■■■■■■

57

7

H

2

O adsorption cartridge (6.2837.010)

For removing the H

2

O from the aspirated air.

Filled with bead desiccant.

8

Adsorption cartridge holder
(6.2057.080)

9

MCS aspirating capillary
Connection to the MCS. Corresponds to
(29-

3

).

Installing the adsorption cartridges

1 Preparing the adsorption cartridge holder

Push the 4 clips (30-3) into the slot of the adsorption cartridge
holder (30-8).

2 Removing the caps

Remove the two sealing caps at the tip of the two cartridges.

In the case of the H

2

O adsorption cartridge, replace the round

sealing cap on the larger end with the star-shaped sealing cap.

3 Connecting the CO

2

adsorption cartridge

Insert the CO

2

adsorption cartridge into the coupling (30-2) on

the end of the MCS aspirating capillary (3-15).

Click the CO

2

adsorption cartridge into the two lower clips (30-3)

of the adsorption cartridge holder (30-8).

4 Connecting the PVC tubing

Insert the adapter (30-5) into the CO

2

adsorption cartridge.

Fasten the PVC tubing (30-6) on the adapter (30-5).

5 Connecting the H

2

O adsorption cartridge

Place the H

2

O adsorption cartridge into the PVC tubing (30-6).

Click the H

2

O adsorption cartridge into the two upper clips (30-3)

of the adsorption cartridge holder (30-8).

6 Placing the adsorption cartridge holder in the instrument

Place the adsorption cartridge holder with cartridges into the
detector chamber of the instrument