Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 22

Advertising
background image

3.2 Initial installation

■■■■■■■■■■■■■■■■■■■■■■

12

■■■■■■■■

881 Compact IC pro – Anion – MCS

Connect one end of the aspirating capillary (6.1803.020) for the
regeneration solution to the tubing olive on the pump tubing
using a short pressure screw (6.2744.070).
Guide the other end of the aspirating capillary out of the instru-
ment through a capillary feed-through, slide it through a bottle
attachment (6.1602.150) and screw the bottle attachment onto
the bottle (6.1608.020) containing the regeneration solution.
Ensure that the end of the aspirating capillary reaches down to
the bottom of the bottle.

Place the pump tubing into a tubing cartridge.

Prepare a second pump tubing for the rinsing solution:

Plug a tubing olive (6.2744.034) onto one end of the pump tub-
ing (6.1826.320).
Plug a pump tubing connection (6.2744.180) onto the other end
of the pump tubing.

Connect one end of the aspirating capillary (6.1803.020) for the
rinsing solution to the tubing olive on the pump tubing using a
short pressure screw (6.2744.070).
Guide the other end of the aspirating capillary out of the instru-
ment through a capillary feed-through, slide it through a bottle
attachment (6.1602.150) and screw the bottle attachment onto
the bottle (6.1608.020) containing the rinsing solution. Ensure
that the end of the aspirating capillary reaches down to the bot-
tom of the bottle.

Place the pump tubing into a tubing cartridge.

Place both tubing cartridges into the peristaltic pump.

6 Connecting the MSM

(see Chapter 3.17, page 51)

Connect the capillary of the MSM labeled with regenerant to the
pump tubing connection of the pump tubing for the regeneration
solution using a short pressure screw (6.2744.070).

Connect the capillary of the MSM labeled with rinsing solution
to the pump tubing connection of the pump tubing for the rinsing
solution using a short pressure screw (6.2744.070).

Guide the two capillaries of the MSM labeled with waste reg.
and waste rins. out of the instrument through a capillary feed
through to a waste container and fasten them there.

7 Connecting the MCS

(see Chapter 3.18, page 54)