3.13 Sample degasser
■■■■■■■■■■■■■■■■■■■■■■
38
■■■■■■■■
881 Compact IC pro – Anion – MCS
1
2
3
4
Figure 19
Sample degasser
1
Sample degasser input
2
Sample degasser output
3
PEEK pressure screw, long
(6.2744.090)
4
Connection capillaries (6.1803.040)
Connecting the sample degasser
1 Remove and keep the threaded stoppers (6.2744.220) from the input
and output of the sample degasser.
2 Connect the end of the sample aspirating capillary (6.1803.040) con-
nected to the injection valve to the output of the sample degasser
(19-2), using a long PEEK pressure screw (19-3).
3 Connect the connection capillary (6.1803.040) to the input of the
sample degasser (19-1), using a long PEEK pressure screw (19-3).
4 Guide the other end of the connection capillary out of the instrument
through a capillary feed-through and connect it with the Sample Pro-
cessor, if applicable.
Caution
If the sample degasser is not used, the input and output must be
sealed with the threaded stoppers (6.2744.220).