Figure 19, Sample degasser – Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 48

Advertising
background image

3.13 Sample degasser

■■■■■■■■■■■■■■■■■■■■■■

38

■■■■■■■■

881 Compact IC pro – Anion – MCS

1

2

3

4

Figure 19

Sample degasser

1

Sample degasser input

2

Sample degasser output

3

PEEK pressure screw, long
(6.2744.090)

4

Connection capillaries (6.1803.040)

Connecting the sample degasser

1 Remove and keep the threaded stoppers (6.2744.220) from the input

and output of the sample degasser.

2 Connect the end of the sample aspirating capillary (6.1803.040) con-

nected to the injection valve to the output of the sample degasser
(19-2), using a long PEEK pressure screw (19-3).

3 Connect the connection capillary (6.1803.040) to the input of the

sample degasser (19-1), using a long PEEK pressure screw (19-3).

4 Guide the other end of the connection capillary out of the instrument

through a capillary feed-through and connect it with the Sample Pro-
cessor, if applicable.

Caution

If the sample degasser is not used, the input and output must be
sealed with the threaded stoppers (6.2744.220).