Metrohm 881 Compact IC pro – Anion – MCS User Manual
Page 72
3.21 Separation column
■■■■■■■■■■■■■■■■■■■■■■
62
■■■■■■■■
881 Compact IC pro – Anion – MCS
2 Rinsing the separation column
■
Place beaker under the outlet end of the separation column.
■
Set the flow rate of the high pressure pump according to the data
given in the leaflet of the separation column.
■
Start the high pressure pump and rinse the separation column
approx. 10 minutes with eluent.
■
Switch off the high pressure pump again.
3 Mounting the separation column
■
Fasten the column output capillary to the output of the separation
column using a PEEK pressure screw (6.2744.070).
■
Hang separation column with chip into the column holder.
Note
The iColumns are equipped with a chip on which their operating data is
saved. The chip has to be hooked into the chip holder provided for this
so that the column recognition can function.