3 selecting the sample loop, 15 column heater, Selecting the sample loop – Metrohm 881 Compact IC pro – Anion – MCS User Manual

Page 51

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Installation

881 Compact IC pro – Anion – MCS

■■■■■■■■

41

1

Eluent input
Capillary coming from the high pressure
pump.

2

Eluent output
Capillary to the column.

3

Sample input
Sample aspirating capillary.

4

Sample output
Capillary to waste container.

5

Sample loop

Position A

In the position FILL, the sample solution flows
through the sample loop to the waste container.
The eluent flows directly to the separation col-
umn at the same time.

Position B

In the position INJECT, the eluent flows through
the sample loop to the separation column. If
sample solution is to be found in the sample
loop at the time of the valve switchover, then
this will be conveyed along with the eluent, thus
making its way to the separation column. The
flow in the sample path is either stopped or the
sample flows directly to the waste container.

3.14.3

Selecting the sample loop

The amount of sample solution injected depends on the volume of the
sample loop. The choice is made on the basis of the application. The fol-
lowing sample loops are normally used:

Cation determination

10 µL

Anion determination with suppression

20 µL

Anion determination without suppression

100 µL

3.15

Column heater

The perfect isolation of the column chamber ensures thermally stable con-
ditions for the separation column. The temperature of the column heater
can be set in the software.

Advertising