7 connection, 8 connection, 5 12 connection – Metrohm 828 IC Dual Suppressor User Manual

Page 13

Advertising
background image

1.3 Parts and controls

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

5

7 To Detector

Inlet connection to detector,

connection for 6.1831.010 PEEK

Capillary with 6.2744.010 PEEK

Pressure screw

10

Flow Cathode

For a visual check of the cathode

eluate. The liquid from the cathode

passes through the FEP tubing before

entering the degassing unit. In normal

operation hydrogen bubbles should be

seen regularly.

8 From Column

Outlet connection from column,

connection for 6.1831.010 PEEK

Capillary with 6.2744.010 PEEK

Pressure screw

11

Detector Flow

The flow to the detector can be

regulated with a screwdriver.

9 Flow Anode

For a visual check of the anode

eluate. The liquid from the anode

passes through the FEP tubing before

entering the degassing unit. In normal

operation oxygen bubbles should be

seen regularly.

12 Waste

Outlet to waste container, connection

for 6.1803.020 PTFE Tubing with

6.2744.010 PEEK Pressure screw