Mains cable and mains connection, Switching the instrument on/off – Metrohm 828 IC Dual Suppressor User Manual
Page 21
2.3 Mains connection
828 IC Dual Suppressor/ 8.828.1003 Instructions for Use
13
2.3.3 Mains cable and mains connection
Mains cable
The instrument is supplied with one of the following mains cables
• 6.2122.020 with SEV 12 plug (Switzerland, …)
• 6.2122.040 with CEE(7), VII plug (Germany, …)
• 6.2133.070 with NEMA 5-15 plug (USA, …)
which has three wires and is fitted with a plug with a grounding pin. If a
different plug has to be used then the yellow/green wire (IEC standard)
must be connected to the grounding pin (Protection class I).
Any interruption to the grounding inside or outside the instrument can
represent a hazard!
Mains connection
Insert the mains cable into mains connector 16 of the 828 IC Dual
Suppressor.
2.3.4 Switching the instrument on/off
The 828 IC Dual Suppressor is switched on and off with mains switch
16. When the instrument is switched on the "Power" 6 LED lights up.