5.5 Unplanned maintenance work

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

37

Figure 9:

Opened IC Dual Suppressor

4 Replace the suppressor cell

• Reconnect the capillaries to the appropriate connections of the

new 6.2838.000 Suppressor cell 26.

The electrode connections are kept waterproof by flat seals.

Tighten them up by hand and then use the pliers to tighten

them by a further 1/8 turn.

The suppressor cell connections and capillaries are marked;

the suppressor cell position and the assignment of the

capillaries is shown in Figure 10.

• Press the connected-up suppressor cell back into the

electrode clamp holders.

Take care that you do not touch the suppressor cell electrodes with

your bare fingers as this could cause contact problems. If you do

accidentally touch the electrodes then they must be degreased with a

little ethanol.

5 Reassemble the instrument and connect it up

• Reattach housing cover 25 by carrying out the steps given in

item 2 in the reverse order.

• Connect the Dual Suppressor to the mains supply and switch it

on. If you have removed the suppressor from your system then

reinstall it as described in Section 2.