2.3 Mains connection

Please observe the following rules when connecting the instrument to

the electricity supply. If the instrument is operated with an incorrectly

set mains voltage and/or an incorrect mains fuse then it represents a

fire hazard!

2.3.1 Mains voltage

All power supplies from 100 V to 240 V with 50/60 Hz can be connected

to the mains input of the 828 IC Dual Suppressor. Adaptation to the

mains supply used is carried out automatically.

2.3.2 Fuse

A 1.0 A fuse (slow blow) is contained in the fuse holder of the 828 IC

Dual Suppressor.

Make sure that the instrument is never operated with a different type of

fuse as otherwise it represents a fire hazard!

Changing the fuse

1 Pull out mains cable

Switch off the instrument and remove the mains cable from mains

supply connection 16 of the 828 IC Dual Suppressor.

2 Remove the fuse holder

Use a screwdriver to loosen the fuse holder beside the mains

supply connection (see Figure 3) and remove it completely.

3 Changing the fuse

Change the fuse if necessary and replace it in the fuse holder.

Use only a fuse with the following specifications.

1.0 A fuse (slow blow)

Metrohm No. U.600.0016

4 Re-insert fuse holder