3 Operation

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

26

3.6 Routine operation

3.6.1 Manual operation

Start sequence

1 Installation

Install the 828 IC Dual Suppressor as described in Section 2.

• The remote connection from the suppressor to the 762 IC

Interface (6.2128.180 Cable, Figure 5) or to the 761 Compact

IC (6.2143.200 Cable, Figure 7) are not necessary for manual

operation.

• The IC Dual Suppressor must be switched to manual opera-

tion, remote switch 15 = OFF.

2 Switch on suppressor

• Switch on the instrument with mains switch 16.

The vacuum pump starts up and the LED "Vacuum" 5 will light

up green as soon as the set working pressure is reached.

Wait until the LED "Vacuum" 5 lights up green before continuing with

step 3.

3 Switch on pump

• Switch on the high pressure pump.

After a short time the pump will build up pressure and, when

the value set for the pressure sensor has been reached (see

page 22), current is applied to the suppressor cell.

The LED "Current" 4 lights up green. This process takes about

1 minute, but this time may vary depending on the set flow rate

and the backpressure of the system. When the cell voltage lies

within the given range then LED "Voltage" 3, followed by LED

"Ready" 1, will light up green.

4 Equilibrate

• Wait 20-30 minutes until the 828 IC Dual Suppressor is

equilibrated.
The suppressor is fully equilibrated when the background con-

ductivity remains constant and a stable baseline is obtained.

When you change the eluent you should first rinse the separation

column without connecting the 828 IC Dual Suppressor. Pump the

eluent through the system for about 5 min.