4 Troubleshooting - Problems

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

30

4

Troubleshooting - Problems

4.1 Remedying faults and problems

If difficulties occur during analyses with your IC system then it is best to

search for their causes in the following sequence: column pump

eluent IC system. In the Instructions for Use of your MIC 10

Modular IC system or 761 Compact IC you will find an overview of

possible faults together with their causes and remedies.

In addition to these general problems, the following section covers

those problems which could arise from the use of the 828 IC Dual

Suppressor.

4.2 Chromatography problems

Problem

Cause

Remedy

Poor peak shape

• Dead volume in system.






• Weak eluent, unsuitable

gradient program.

• Space inside the suppressor

cell.

• Check that the capillary connections are

tight. Keep the length of the PEEK
capillaries between the column and IC
Dual Suppressor and between the IC Dual
Suppressor and detector as short as
possible and make sure that a capillary

with the correct inner diameter is used
(6.1831.010).

• Make up a new eluent, correct the

gradient.

• Exchange suppressor cell, see Section

5.5.2.

Poor retention time
reproducibility

• Leak in the instrument.

• Flow rate ratio not correct.

• Blockage in flow path.

• High pressure in system.

• Check whether liquid is escaping from

drain 20, replace leaking connections.

• Check the flow rate ratio, Section 3.5.

• See Section 4.3.

• See Section 4.3.