3.6 Routine operation

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

29

When you change the eluent you should first rinse the separation

column without connecting the 828 IC Dual Suppressor. Pump the

eluent through the system for about 5 min.
If you change the eluent type, particularly when changing from a

strong to a weak eluent, then it may take longer to equilibrate the

suppressor.

4 The system is now ready to carry out an analysis


Stop sequence

1 Switch off system

• Shut down the system with Control/Shutdown hardware. On

the 828 IC Dual Suppressor the current to the suppressor cell

and the vacuum pump are switched off.

If the IC Dual Suppressor is to be used again within a few days

then no further measures are necessary.

The following steps are only necessary if the suppressor will not be

used for more than one week.

2 Rinse suppressor

• Before switch-off let the system run with pure water for 15 min

(1 mL/min); the suppressor must be working.

3 Switch off suppressor

• Shut down the system with Control/Shutdown hardware. On

the 828 IC Dual Suppressor the current to the suppressor cell

and the vacuum pump is switched off.

• Switch off the suppressor at mains switch 16.

Seal the inlet of the CO

2

absorber cartridge when the 828 IC Dual

Suppressor is not to be used for a longer period.

4 Storage

These measures are only necessary when the IC Dual Suppressor

will not be used for a long time.

• Remove the connections from the following inlets and outlets:

"To Detector" 7, "From Column" 8 and "To Waste" 12 and

seal them with the original closures supplied with the

instrument.