2 Installation
828 IC Dual Suppressor/ 8.828.1003 Instructions for Use
18
• Double-click on the instrument symbol of the 762 IC Interface to
open its configuration menu and select the file card Events setup.
• The value of the event line, via which the Dual Suppressor will be
controlled, must be set to 1. In this example this is Event Line 1, see
• Set Event Line 1 to the value 1 and confirm with OK.
The 828 IC Dual Suppressor is now included in the system and will be
controlled by the Metrodata IC Net 2.1 software. It is started together
with the other hardware of the system "Gradient.smt" with SYSTEM
/Control/Startup hardware and shut down with SYSTEM
/Control/Shutdown hardware.
In Metrodata IC Net software versions with version numbers higher
than 2.1 the 828 IC Dual Suppressor will be preconfigured in the
software just like all other Metrohm IC instruments. It can then be
immediately installed in the New System Wizard. A detailed
description can be found in the corresponding Instructions for Use of
the software.