3.6 Routine operation

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

27

If you change the eluent type, particularly when changing from a

strong to a weak eluent, then it may take longer to equilibrate the

suppressor.

5 The system is now ready to carry out an analysis


Stop sequence

1 Switch off pump

Step 2 is only necessary when the suppressor will not be used for

more than one week; otherwise continue with step 3.

2 Rinse suppressor

• Leave the suppressor switched on and rinse it with pure water

for 15 min (1 mL/min).

3 Switch off suppressor

• Switch off the suppressor at mains switch 16.

You can also leave the 828 IC Dual Suppressor switched on when it is

not being used; this will not damage the instrument. However, when it

is switched on the vacuum pump operates continuously. This is why

we recommend that the 828 IC Dual Suppressor is always switched

off when it is not being used in order not to unnecessarily shorten the

working lives of the vacuum pump and CO

2

absorber cartridge.

You should also seal the inlet of the CO

2

absorber cartridge when the

828 IC Dual Suppressor is not to be used for a longer period.

4 Storage

These measures are only necessary when the IC Dual Suppressor

will not be used for a long time.

• Remove the connections from the following inlets and outlets:

"To Detector" 7, "From Column" 8 and "To Waste" 12 and

seal them with the original closures supplied with the

instrument.


3.6.2 Software control

The operation and functions of the two control programs

IC Net 2.1

and

761 PC Software 1.1 are identical in many respects. Where

differences occur these are described separately for the particular

software.