3 Operation

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

24

removed and the equilibrium for the breakdown of the carbonic acid is

displaced even further toward the dissociated products. The CO

2

absorber cartridge is required to remove CO

2

and humidity from the

drawn-in ambient air before it enters the degassing system. Exchanging

the CO

2

absorber cartridge is described in Section 5.4.1.

The vacuum pump works at a constant speed and as soon as the

vacuum reaches the preset working pressure LED "Vacuum" 5 lights up

green.

3.5 Flow rate ratio

Please ensure that the backpressure capillary is mounted at the outlet

of the detector used, see Section 2.4.2, Section 2.5.2; it is required for

the correct ratio between the flow rates.
Check this ratio between the flow rates before starting up the 828 IC

Dual Suppressor for the first time.


The 828 IC Dual Suppressor achieves its optimal performance when

40% of the total flow is eluted to the detector. The 828 IC Dual

Suppressor is preadjusted for a total flow rate of 1 mL/min. At a total

flow rate in the working range of 0.5 - 2.0 mL/min this flow ratio only

changes slightly, but you should nevertheless check it when you alter

the total flow rate.
If more than one detector is used in series downstream the IC Dual

Suppressor then the ratio between the flow rates must be adjusted in

order to compensate for the additional backpressures of the other

detectors.

3.5.1 Setting the flow rates

The IC Dual Suppressor is designed to work at a flow rate to the

detector of 40% ± 5% of a total flow rate of 1 mL/min.
The following instructions describe in detail how you can check the ratio

between the flow rates and alter it if necessary:

1 Preparation

Set the flow rate at the pump to 1 mL/min and wait for 10 minutes.

2 Measurement

Measure the volume V or the weight m of the liquid which elutes

from the backpressure capillary in 3 minutes.