2 Installation

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

20

2 To Detector

• Use another piece of 6.1831.010 PEEK Capillary (i.d. =

0.25 mm, o.d. = 1/16"), further 6.2744.010 Pressure
screws
and a 6.2744.040 Coupling (all included in
standard equipment) to connect the outlet of the 828 IC Dual
Suppressor "To Detector" 7 with the inlet of the conductivity

detector in the 761 Compact IC.

3 Detector outlet

• Mount the 6.1803.090 PTFE Capillary (i.d. = 1/16", o.d. =

0.25 mm, 15 m) as a backpressure capillary between the
detector outlet and the waste container. Use a 6.2744.010
Pressure screw and a 6.2744.040 Coupling (all included in
standard equipment) for this.

The backpressure capillary is required to ensure the correct ratio

between the flow rates at the outlets of the suppressor cell, see

Section 3.5.
Check this ratio the

first time that a system in which the 828 IC Dual

Suppressor has been newly installed is started up; a detailed

description is given in Section 3.5.1.

4 To Waste

• Connect outlet "Waste" 12 of the 828 IC Dual Suppressor to a

waste container. Use the 6.1803.020 PTFE Tubing (i.d. =
0.97 mm, o.d. = 1.57 mm) and a 6.2744.010 Pressure

screw (both included in standard equipment) for this.

PEEK capillaries that are to be provided with new connections must

have a perfect and plane cutting surface. This is best obtained by

using the optionally available

6.2621.080 Capillary cutter .

2.5.3 Settings for control by the 761 PC Software 1.1

On the 828 IC Dual Suppressor switch remote switch 15 to "ON" and

also switch mains switch 16 to "ON". The Dual Suppressor can now

be controlled by an external instrument.

In the 761 PC Software 1.1 you must make the following settings:

• Start the 761 PC Software 1.1 and select your system.