Safety notes, Electrical safety, General safety measures – Metrohm 828 IC Dual Suppressor User Manual

Page 17

Advertising
background image

1.5 Safety notes

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

9

1.5 Safety notes

Warning!

This instrument should only be used in accordance with the

information given in these Instructions for Use.

1.5.1 Electrical safety

Electrical safety when handling the 828 IC Dual Suppressor is

guaranteed within the scope of Standard IEC 61010. However, please

observe the following points:

• Mains connection

The

mains connection and checking the mains fuse must be

carried out according to the information given in Section 2.3.

• Opening the 828 IC Dual Suppressor

If the 828 IC Dual Suppressor is connected to the mains supply then it

must neither be opened nor should any parts be removed from it, as

otherwise the risk of contact with current-carrying assemblies exists.

This is why the instrument should be separated from all voltage

sources before being opened. Please make sure that the

mains

cable is pulled out from mains connector 16!

• Protection against electrostatic charges

Electronic components are sensitive to electrostatic charges and can

be destroyed by a discharge. Before touching any assembly within

the 828 IC Dual Suppressor you should ground yourself and any tool

you are using by grasping a grounded object (e.g. the instrument

housing or a radiator) in order to eliminate any electrostatic charge

which may be present.


1.5.2 General safety measures

• Liquid handling

Check all inlet and outlet tubing for leaks at regular intervals. Observe

the appropriate regulations concerning the handling of flammable

and/or toxic solutions and their disposal.