Connecting the ic dual suppressor, Figure 6, Flow diagram of the mic 10 – Metrohm 828 IC Dual Suppressor User Manual

Page 23

Advertising
background image

2.4 Connecting to a modular gradient IC system

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

15

2.4.2 Connecting the IC Dual Suppressor

The following diagram shows how the system is arranged:

Figure 6:

Flow diagram of the MIC 10

All the components should be connected as shown in Figure 6, which

also provides a flow diagram for the sample.

The installation of the other components is described in detail in the

appropriate Instructions for Use.

The 828 IC Dual Suppressor is inserted between the separation column

and the detector. The following connections must be made:

Please ensure that the liquid connections between the separation

column outlet, the Dual Suppressor and the detector inlet are as short

as possible and use a narrow diameter capillary, e.g. 6.1831.010

PEEK Capillary (included as standard). In this way you minimize the

dead volumes and reduce the peak broadening of the signal.

IC Pumps

IC Detector

IC Dual Suppressor

Sample

Gradient
mixing capillary

Pulsation
dampener

T-connector

Separation
column

Sample
loop

A Trap column

Valve

Inline filter