5.5 Unplanned maintenance work

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

35

4. Pump ultrapure water at 1 mL/min through the system with the

suppressor switched on (mains switch 16 = ON).

When LED "Voltage" 3 lights up green continue for a further 30

min; the suppressor is then regenerated.

If LED "Voltage" 3 turns red again then repeat rinsing step 3.

and continue with 4. If the LED "Voltage" 3 still remains red

then regeneration procedure 2. - 4. must be repeated.

If the suppressor cannot be regenerated then the suppressor

cell must be replaced.

5.5.2 Replacing the suppressor cell


If the suppressor cell can no longer be regenerated then it must be

replaced.

As replacing the suppressor cell requires the instrument to be

opened, this should only be carried out by experienced users. Please

observe the safety information given in Section 1.5.1.

If the 828 IC Dual Suppressor is connected to the mains supply then it

must neither be opened nor should any parts be removed from it, as

otherwise the risk of contact with current-carrying assemblies exists.

This is why the instrument should be separated from all voltage

sources before being opened. Please make sure that the

mains

cable is pulled out from mains connector 16!

Required tools:

1 Phillips screwdriver

1 Pliers

1 Preparation

• Switch off the 828 IC Dual Suppressor at mains switch 16.

• Remove the mains cable from connection 16.