6.4 Validation / GLP

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

45

6.4 Validation / GLP

GLP (Good Laboratory Practice) requires, among other things, that the

precision and correctness of analytical instruments is checked at

regular intervals by using SOPs (Standard Operating Procedures,

SOP). An example of such a standard operating procedure is available

from Metrohm under the title «Application Bulletin No. 277 –

Validation of Metrohm Ion Chromatography Systems by using

Standard Operating Procedures (SOP)». This SOP can be adapted

for your ion chromatography system and used for its validation.
The 828 IC Dual Suppressor must be included as a part of the whole

ion chromatography system, whose most important components

include the pumps, separation columns, detector and evaluation

system, in the all-embracing validation of the whole system.
Please contact your local Metrohm agency in order to receive support

in validating your 828 IC Dual Suppressor. It can also provide you with

validation documentation which will help you to carry out your

installation qualification (IQ) and operational qualification (OQ).
Further information about QA, GLP and validation can also be found in

the brochure «Quality management with Metrohm» which is also

obtainable from your local Metrohm agency.
Checking the electronic and mechanical assemblies of Metrohm

instruments can and should be undertaken within the framework of

regular servicing by Metrohm technicians (see Section 5.2). All Metrohm

instruments are equipped with start-up check routines which check that

the relevant assemblies are functioning perfectly when the instrument is

switched on. If no error message appears it can be assumed that the

instrument is functioning properly.
The 828 IC Dual Suppressor also contains a built-in diagnosis program

which allows the service technicians to check the functioning of

particular assemblies should faults or malfunctions occur and to

localize them.