Metrohm 828 IC Dual Suppressor User Manual

Page 39

Advertising
background image

4.2 Chromatography problems

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

31

Noisy or unstable baseline • Suppressor cell is not yet

properly equilibrated with the
eluent.

• CO

2

absorber cartridge is

exhausted.

• Fraction of volatile organic

modifiers in eluent is too high.

• Flow rate ratio not correct.

• Leak in the instrument.

• Blockage in flow path.

• Let the system run and wait until the

suppressor is equilibrated.

• Exchange the 6.2827.000 CO

2

absorber

cartridge, see Section 5.4.1.

• Make up a new eluent, refer to Section

3.2.

• Check the flow rate ratio, Section 3.5.

• Check whether liquid is escaping from

drain 20, replace leaking connections.

• See Section 4.3.

Background conductivity
too high

• Eluent concentration incorrect.

• Eluent contaminated.

• CO

2

absorber cartridge is

exhausted.

• Flow rate ratio not correct.

• Suppressor cell is exhausted.

• Blockage in flow path.

• Make up a new eluent, take the capacity

of the suppressor into account, see

Section 3.3.

• Make up a new eluent, see Section 3.2.

• Exchange the 6.2827.000 CO

2

absorber

cartridge, see Section 5.4.1.

• Check the flow rate ratio, Section 3.5.

• Regenerate the cell, proceed as

described in Section 5.5.1.

• See Section 4.3.