2.5 Connecting to the 761 Compact IC

828 IC Dual Suppressor/ 8.828.1003 Instructions for Use

19

2.5 Connecting to the 761 Compact IC

The installation of the 828 IC Dual Suppressor with the 761 Compact IC

and control via the «761 PC Software 1.1» software is described

below.

2.5.1 Electrical connection

The electrical connections of the system consisting of the 761 Compact

IC and the 828 IC Dual Suppressor are made according to Figure 7:

Figure 7:

Connecting the 828 IC Dual Suppressor
to the 761 Compact IC

2.5.2 Connecting the IC Dual Suppressor

Please ensure that the liquid connections between the separation

column outlet, the Dual Suppressor and the detector inlet are as short

as possible and use a narrow diameter capillary, e.g. 6.1831.010

PEEK Capillary (included as standard). In this way you minimize the

dead volumes and reduce the peak broadening of the signal.

1 From Column

• Connect the outlet of the separation column in the 761

Compact IC to the inlet of the 828 IC Dual Suppressor "From
Column
" 8. Use a 6.1831.010 PEEK Capillary (i.d. = 0.25
mm, o.d. = 1/16") and a 6.2744.010 Pressure screw (both

included as standard) for this.

761

828

6.2143.200

6.2134.100

PC

Advertising