2.2 Mains connection

817 Bioscan

13

4 Insert fuse

Change fuses if necessary and reinsert in fuse holder.

5 Install fuse holder

Depending on the necessary mains voltage, insert fuse holder in
the 817 Bioscan so that the corresponding mains voltage range
can be read normally and the adjacent white arrow points to the
white bar imprinted below the fuse holder (see Fig. 4).

18

220 - 240 V

100 - 120

V

17

16

220 - 240

V

100 - 120 V

Fig. 4: Setting the mains voltage

2.2.2 Fuses

Two fuses 3.15 A/slow-blow for 100

…120 V or 1.6 A/slow-blow for

220

…240 V are installed in fuse holder 16 of the 817 Bioscan as stan-

dard.

Ensure that the instrument is never put into operation with fuses of a
different type as this could cause a fire!

For checking or changing fuses, proceed as described in section 2.2.1.