3 connecting a sample changer, 2 installation – Metrohm 817 Bioscan User Manual

Page 30

Advertising
background image

2 Installation

817 Bioscan

22

2.4.3

Connecting a sample changer

If the MIC-8 system described above is to be operated with a Metrohm
sample changer then the following connections and settings must be
made:

709

812

817

6.2128.100

1

2

3 4 5

6 7 8 9

PC

6.2125.160

Gr

ou

nd

Fi

ll

In

ject

AU

X

1

AU

X

2

1

2

10 11

12

CE

L

L

O

F

F

CE

L

L

O

N

+

24

V

GR

OU

N

D

766/788/813

6.2141.140

6.2141.140

Fig. 8: Connecting a simple MIC 8 system including a sample

changer

The IC Sample Processor 766, the IC Filtration Sample Processor
788
or the Compact Autosampler 813 can be used as the sample
changer. In this combination of the MIC-8 system the 817 Bioscan is
operated as the "Master". The start impulse for the instrument methods
to the connected sample changer is transferred via the remote connec-
tion (6.2141.140, optional). You can then select any instrument method
at the sample changer that is described as being for control with the PC
as master. Information about this is given in the corresponding ‘Instruc-
tions for Use’.

If a second COM port on the PC or a 762 IC Interface is available for an
RS232 connection then the 766 IC Sample Processor or the 788 IC Fil-
tration Sample Processor can be connected up via this connection. In
this way you can generate your own time programs in IC Net 2.1 for
these sample changers. Details are given in the ‘Instructions for Use’ for
the particular instrument.