Metrohm 817 Bioscan User Manual

Page 31

Advertising
background image

2.4 System for sugar analysis – MIC-8

817 Bioscan

23

Creating the time program

In order to control the 766 IC Sample Processor and the 788 IC Fil-
tration Processor
the following program is created in the Bioscan:

In this way shortly after the injection valve has been switched to 'Fill' at
the 812 Valve unit, the first remote line 'AUX 1' (see Fig. 6) will be
switched briefly to 'ON'. This signal will be scanned at the start of the
suitable sample changer programs.

Terminate the program entry by activating the program with

Enable

and

store it in the system with

Save

.

The control of the 813 Compact Autosampler is carried out similarly.
The only thing to be done here is to set the remote lines to the required
times as described above for each of the three working steps: auto-
sampler start, pump start and pump stop in order for a brief impulse to
be transferred.

In order for the connection Autosampler – 817 Bioscan to function
properly the program must first be started at the sample changer and
then the "Sample queue" in IC Net 2.1.