2.3 Capillary connections

817 Bioscan

17

2.3.5

Connection of the flowcell

The Flowcell (6.1254.010) is a part of the standard equipment of the
817 Bioscan. It is already mounted and ready to use.

35

36 37

38

39

40

41

Fig. 6: Connections at flowcell

35

Check marks
For correct orientation of the top and
bottom part of the flow-cell

39

Auxiliary electrode connection
for blue electrode cable

20

36

Mounting screws (4x)

40

Reference electrode connection
for black electrode cable

24

37

Outlet for eluent
Connection of PEEK capillary by
6.2744.130 KELF pressure screw .
PEEK pressure screws may damage
the threads of the cell.

41

Inlet for eluent
Connection of PEEK capillary by
6.2744.130 KELF pressure screw.
PEEK pressure screws may damage
the threads of the cell.

38

Working electrode connection
for red electrode cable

19

1 Before starting up check all screws 36

• Use the Allen key provided to loosen all four of the 36 screws

by one rotation.

• Then slightly tighten the screws again, e.g. by holding the

Allen key between the thumb and index finger only.

• Finally tighten all four screws by a quarter rotation.
• Always tighten opposing screws, i.e. tighten one screw and

then the opposite screw.

• Avoid overtightening the screws as this could damage the

6.1254.020 Distance piece!