6.1 Technical data

817 Bioscan

67

Signal output

IE convertor

Azero
Offset

ADC

CPU

DAC

REC

INT

Icell

RS232

Cell

I = I

+ I

+ I

cell

ADC

Azero

offset

RS232 INT

REC

output
signal

+/- 21 µA
( with comp.)

+/- 10 V

+/- 1V

Max. range

5 nA, 50 nA, 5 µA *

idem

idem

Max. comp.

16 nA, 160 nA, 16 µA idem

idem

Max. signal

21 nA, 210 nA, 21 µA idem

idem

Resolution

1 pA, 10 pA, 1 nA

Unlimited,
purely analogue
(only limited by
acquisition software)

12 bits (0.5 mV)

Auto zero

Resets zero point of
full scale window **

Set to 0 V

Set to 0 V

Offset

Resets full scale
window to % of full
scale range ***

Set relatively to
subrange setting

Set as % of full scale
output

Filter
(DC only)

Y N Y

* In RS232 data acquisition use always the maximum range settings. Minor

range settings do not amplify the signal, however they limit the full scale.
Minor range settings affect (amplify) the REC output only.

** If a 5 µA range setting is used, the full scale window is –5..+5 µA.After auto

zero at a background current of 1 µA and, this full scale window is –4..+6 µA.

*** If a 5 µA range setting is used, the full scale window is –5..+5 µA. Applying –

50% offset, results in a full scale window of –2.5..+7.5 µA.