3 Operation

817 Bioscan

46

List of program instructions

The following program instructions can be added to the time
program on the

Program

subpage:

Instruction Parameter

entry Meaning

Mode

DC, Pulse, Scan

Select detector mode.

Cell On,

Off

Switch cell on/off.

Outputs

0, 1, *

Set remote lines Aux1, Aux2, Cont1
and Cont2 to desired values. For entry
of the first value, enter

1

,

0

or

*

. For

entry of the other values, move the
cursor in front of the value to be
changed and enter

1

,

0

or

*

.

Range

pA: 10 pA – 5 nA
nA: 0.1 nA – 50 nA
µA: 10 nA – 5 µA

Set measuring range.

Offset

-50, -40, -30,…50 %

Set offset for measuring signal.

Valve Fill,

Inject

Switch connected injection valve.

Oven

off, 15 ... 60 °C

Set column oven temperature.

Autozero

Trigger analog output signal auto
zero
.

DC_E -2 ... +2 V

Set

measuring potential (DC).

Filter

0.1, 0.2, 0.5 ... 5 s

Set electronic damping for the
analog output signal.

Scan_E1

-2 ... +2 V

Set sweep start potential E1.

Scan_E2

-2 ... +2 V

Set sweep end potential E2.

Scan_Cycle half,

full

Select scan cycle type.

Scan_Rate

1, 2, 5, … 50 mV/s

Set scan rate.