6 connecting the waste tubing – Metrohm 817 Bioscan User Manual

Page 26

Advertising
background image

2 Installation

817 Bioscan

18

2 Connect the electrode cable

• The included 6.2156.000 Cable for connecting the measur-

ing cell is attached to connections

38, 39 and 40 of the

measuring cell using the three connectors

19, 20 and 24

(see Fig. 6).

• The other end of the cable is inserted into connection '1' of

the 817 Bioscan 817 with connector

27 (see Fig 3).

3 Connect the column

• Screw column connection capillary 21 of the column on to

eluent inlet

41 of the flow-through measuring cell using the

6.2744.130 Pressure screw.

4 Connect the waste container

• In order to ensure an adequate backpressure, screw a PEEK

capillary (approx. 1 m) on to eluent outlet

37 of the measuring

cell.

• Lead the end through opening 3 into a sufficiently large waste

container and attach it there.

5 Mount the measuring cell

• Attach the flow-through measuring cell to the holder 22 817

Bioscan which is provided for it (see Fig. 3).

• Rotate the holder and the cell so that the cell outlet is located

as high as possible. In this way it is easier for any air bubbles
that may occur to escape from the cell.

Never switch on the flow-through cell when this is

(a) not being rinsed through with a conducting eluent at the

time,

(b) not completely connected up, or

(c) when its outside is moist so that so a short circuit could

occur between the connections of working electrode

38

and auxiliary electrode

39!

2.3.6

Connecting the waste tubing

The 817 Bioscan has an outlet

13 on its rear panel for discharged

liquids to which the waste tubing can be attached. Proceed as follows:

1 Connect the waste tubing

• Attach matching tubing (e.g. Silicone tubing, optional) into

outlet

13 (see Fig. 2).

2 Lead the waste tubing to a drain

• Place the other end of the waste tubing in a drain and attach

it there.