Page 49

Advertising
background image

3.2 Operation using «IC Net 2.1»

817 Bioscan

41

Pulse mode

The

Pulse mode

subtab of the

Method parameters

tab on the

817 Bioscan

window is used to set the parameters for the

pulse mode. This window is only available if

Mode

has been

set to

Pulse

on the

Manual

tab.

Method data

Cell

Definition of three different working poten-
tials that are applied to the amperometric
detector cell cyclically. The set potentials
are graphically displayed if the

Method

op-

tion is enabled above the graphic window.

E1, E2, E3

Potentials.
Range:

-2 ... +2 V

.

t1, t2, t3

Time duration to apply the potentials.
Range:

0.01 ... 2 s

Sample time

Measuring time at the end of working po-

tential

E1

.

Selection:

20, 40, 60, 80, 100 ms

Oven

Switches column oven on/off and sets op-
erating temperature.
Range:

15 ... 60 °C

(> room temp. + 10 °C)

Range

Selection of measuring range:

pA

10 ... 5000 pA

.

nA

0.1 ... 50 nA

.

µA

0.01 ... 5 µA

.

In

the

Pulse mode,

always the

µA

range

should be selected.