Metrohm 817 Bioscan User Manual

Page 53

Advertising
background image

3.2 Operation using «IC Net 2.1»

817 Bioscan

45

Caption

Possibility of entering a title that is written
at the head of the scan mode graphics
window.

<Add date & time>

The current date and time are added to

the title defined in the

Caption

field.

State

Display of current state.

<Start>

Starts scan mode sweep.

<Stop>

Stops scan mode sweep.

<Copy to clipboard>

Copies the content of the scan mode

graphics window to the clipboard.

Program

On the

Program

subpage, program steps including time,

program instruction and parameter can be entered.

First column

Time at which program instruction is
applied.
Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is applied to-

gether with the last instruction with time
entry.

Second column

Program instruction (see below).

Third column

Parameter for program instruction (see
below).

ENABLED

Enable program start (a disabled pro-
gram is not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages
are displayed if program is wrong).