3.2 Operation using «IC Net 2.1»
817 Bioscan
45
Caption
Possibility of entering a title that is written
at the head of the scan mode graphics
window.
<Add date & time>
The current date and time are added to
the title defined in the
Caption
field.
State
Display of current state.
<Start>
Starts scan mode sweep.
<Stop>
Stops scan mode sweep.
<Copy to clipboard>
Copies the content of the scan mode
graphics window to the clipboard.
Program
On the
Program
subpage, program steps including time,
program instruction and parameter can be entered.
First column
Time at which program instruction is
applied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruction is applied to-
gether with the last instruction with time
entry.
Second column
Program instruction (see below).
Third column
Parameter for program instruction (see
below).
ENABLED
Enable program start (a disabled pro-
gram is not started).
<Add>
Add new program instruction.
<Delete>
Delete selected program instruction.
<Verify>
Test the time program (error messages
are displayed if program is wrong).