2 settings in «ic net 2.1 – Metrohm 817 Bioscan User Manual

Page 29

Advertising
background image

2.4 System for sugar analysis – MIC-8

817 Bioscan

21

2.4.2

Settings in «IC Net 2.1»

1 System

As in this configuration only the 817 Bioscan is to be directly
controlled by IC Net then only this instrument is to be installed as
the detector with the help of the system assistant

New system

wizard

.

2 Method

You should first select the standard method supplied for use
with your particular column (e.g. for Metrosep Carb 1:

CARB1_250.mtw

). You can save any alterations under a different

file name.

3 Data source

Select:

Cell( 817 BioScan)

as the signal source.

4 Save system

Then save the system (e.g.

817_MIC-8.smt

).

The following system window should now appear:

5 Connect system and workplace

Under

SYSTEM / Control / Connect to workplace

connect the

new system with the workplace.

At the same time the symbol

will appear in the symbol

bar.

6 Check measuring cell

If the 817 Bioscan is switched on without the flow-through cell
being constantly rinsed by a conducting eluent (e.g. 0.1 M
NaOH) then the measuring cell must not be switched on!

You should now check this in the instrument window of the 817
Bioscan (accessed with a double-click with the mouse on in-
strument symbol) under

Manual / Cell

.

7 Check start mode

Make sure that in this system the 817 Bioscan is standing under

Start with determination

under

Setup / Start mode

.

.