HS 2000 User Manual

Part 3 - Calibration

HB-0846-03

3-13

3.5.1

Sea Temperature Offset Calibration

To calibrate SEA TEMP C or SEA TEMP F proceed as follows:

(1) Select SEA TEMP C on upper half on FFD display.
(2) Press Scroll Down until the lower text shows
CALIBRATE flashing.
(3) Press Enter, the lower text now shows CAL VAL 1
flashing.
(4) Press Scroll Down, the lower text now shows CAL
VAL 2 flashing.
(5) Press Enter, the lower text now shows OFFSET C.
(6) Press Enter, the lower text now shows OFFSET C
flashing and by use of Scroll Up/Down the temperature
should be changed to the reference value.
(7) Press Enter, the offset value is accepted and the
upper display will show the adjusted measured
temperature.

Similarly the above calibration can be carried out if SEA TEMP F
is initially selected.
The offset value is automatically converted so that both degrees C
and F are adjusted correctly.