Calibration

The

820i

can be calibrated using the front panel, serial commands, or

Revolution

®. Each method consists of the

following steps:

Zero calibration

Entering the test weight value

Span calibration

Optional five-point linearization

Optional rezero calibration for test weights using hooks or chains

The following sections describe the calibration procedure for each of the calibration methods. See the

iQUBE

Installation Manual, PN 77224, for information about configuring

iQUBE

-attached scales.

SCALES SERIAL FEATURE

PFORMT

SETPTS

DIG

I/O

ALGOUT

VERS

SCALE 1

CALIBR

WZERO

WVAL

WSPAN

WLIN

REZERO

POINT 4

POINT 5

POINT 1

POINT 2

POINT 3

DISPLAY

Figure 4-1. Calibration (CALIBR) Submenu

4.1

Gravity Compensation

Gravity compensation for latitude and elevation is available for the

820i

. To calibrate with gravity compensation,

the LOCALE parameter under the FEATURE menu must be set ON, and the LATUDE (latitude) and ELEVAT

(elevation, in meters) parameters set before calibrating the indicator (see Figure

3-11 on page

35).

If the indicator is later installed at a different location, gravity compensation can be applied to a pre-calibrated

indicator by adjusting the LATUDE and ELEVAT parameters.

4.2

Front Panel Calibration

The CALIBR submenu (under the SCALES menu,

see Figure

4-2) is used to calibrate the

820i

.

05/24/2006

02:57PM

Calibrate

WZERO

WVAL

REZERO

1291

Deadload Calibration

+/- Millivolts

Zero Count =>

Last Zero

Temp Zero

Figure 4-2. WZERO Calibration Display

The zero, span, and linear calibration point displays

provide a set of softkeys used specifically for

calibration procedures:

+/–

Toggles to allow entry of negative or

positive values

Last Zero

Recalls the last established zero value

to allow calibration without removing

tests weights or product from scale.

Calibrate

Performs calibration for the selected

point

Temp Zero

Temporarily zeroes the displayed

weight of a non-empty scale. After span

calibration, the difference between the

temp zero and the previously calibrated

zero value is used as an offset.

Millivolts

(or

Counts

)

Toggles between display of captured

A/D counts and captured millivolts

values; allows entry of calibration

values in mV or counts