4 revolution calibration, Revolution calibration – Rice Lake 820i Programmable Indicator/Controller - Installation Manual User Manual

Page 54

Advertising
background image

50

820i

Installation Manual

5. Up to five linearization points can be calibrated

between the zero and span calibration values.

Use the following commands to set and

calibrate a single linearization point:

SC.WLIN#n.V1=vvvvv<CR>
SC.WLIN#n.C1<CR>

The SC.WLIN#n.V1 command sets the test

weight value (vvvvv) for linearization point 1.

The SC.WLIN#n.C1 command calibrates the

point. Repeat using the SC.WLIN#n.Vx and

SC.WLIN#n.Cx commands as required for

additional linearization points.

6. To remove an offset value, clear all weight

from the scale, including hooks or chains used

t o h a n g t e s t w e i g h t s , t h e n s e n d t h e

SC.REZERO#n serial command.

7. Send the KSAVE serial command to save the

calibration changes; send KEXIT to exit setup

mode.

4.4

Revolution Calibration

The

Revolution

Calibration Wizard provides

step-by-step scale calibration. With the

820i

connected to the PC, select the Calibration Wizard

from the Tools menu on the

Revolution

Scales display,

then follow the steps listed below to calibrate the

scale. See Section 5.0 on page 51 for a general

overview of the

Revolution

utility.

1. On the first Calibration Wizard display (see

F i g u r e 4 - 3 ) , s e l e c t w h e t h e r y o u a r e

performing a standard (zero and span)

calibration or a multi-point linear calibration.

Click the

Next

button to continue.

Figure 4-3. Revolution Calibration Wizard Display

2. Enter the test weight used to calibrate the

scale. If chains or hooks are used to hold the

weights, check the box below the test weight

value entry. This adds a rezero step to the

calibration sequence.

3. Remove all weights from the scale. If chains

or hooks are used, place them on the scale.

Press the

Calibrate Zero

button to perform the

zero calibration. A message box appears when

the process is complete.

4. Apply the test weights to the scale. Press the

Calibrate Span

button to perform the span

calibration. A message box appears when the

process is complete.

5. If the option for chains or hooks was selected

in step 2, the Rezero display is shown.

Remove all weights from the scale, including

chains or hooks. Press the

Re-Zero

button to

calibrate the zero offset.

6. If performing a multi-point linear calibration,

up to five other calibration weight values can

be entered on the display. The weights must

be in ascending order and must not include

zero or the span weight. Enter the weight

values and click the

Go

button to calibrate

each point.

7. Review the new calibration values then click

Finish

to close the Calibration Wizard. To

restore the current calibration values, click

Cancel

.

Advertising