10.0 Appendix

10.1 Two-Channel IQube Diagnostics

The two-channel iQube provides a number of diagnostic tests. See the table below for a list of diagnostic error
messages. It also describes the tests that can be enabled or disabled in the diagnostics of the two-channel iQube.

Table 10-1. Two-Channel iQube Diagnostic Settings

Test

Definitiion

Description

920i Default

Revolution III/

two-channel Default

Zero
Reference

Which cell is out of tolerance
at zero

Threshold: the upper limit of the zero reading
of the cells

0.125mV

0.200 mV

Percent threshold: the percentage of zero
reading of adjacent cells

25%

20%

Delay: time to wait until the next testing
process

900 seconds

900 seconds

Load Cell
Drift

Checks for load cell outputs
that have changed over time
with load applied

Limit

0.025mV

0.025mV

Percent limit

10%

10%

Time

30 seconds

30 seconds

Load: minimum load needed to initiate the
test

1000 lb

1000 lb

Peak to
Peak Limit

Check for noisy load cell
output not attributed to
environmental conditions

Limit

0.125mV

0.250mv

10.1.1 Diagnostic Setup

Parameters associated with each of the two-channel iQube’s diagnostics test can be set using the 920i menus, iRev,
or Revolution III.

When using the two-channel iQube with the 920i indicator, a Diagnostics softkey can be configured to display
diagnostic information when in weigh mode. Use the FEATURE menu, SOFTKEYS submenu, to select the
diagnostics softkey.