4 characteristics in continuous operation – ALTANA LCS III User Manual

Page 15

Advertising
background image

14

Start Up

The Main Menu is displayed when diagnostics are completed.

See section

5.1.3 Main Menu on page 16

for a detailed description.

4.4 Characteristics in continuous operation

After a continuous operating time of more than 24 hour without
switching off the instrument we recommend to perform a new
calibration with a distilled water cuvette/sample cell.

After a continuous operating time of more than 7 days without
switching off the instrument we recommend to switch off and on the
instrument to perform a new system check with filter adjustment
and lambda-calibration.