59

Standard Operations

λ / Absorbance formula:

1.

Press Absorbance Formula.

2.

The formula selected in the top key determines the number of
wavelength and coefficent keys that will appear below. To
change the absorbance formula, press the top key, select a
formula from the displayed list and press OK. When a new
formula is selected, the number of variables below changes to
match.

The following formulas are available:
K

1

A

1

+ K

2

A

2

K

1

A

1

+ K

2

A

2

+ K

3

A

3

K

1

A

1

+ K

2

A

2

+ K

3

A

3

+ K

4

A

4

K

1

A

1

/ K

2

A

2

(K

1

A

1

+ K

2

A

2

) / K

3

A

3

(K

1

A

1

+ K

2

A

2

) / (K

3

A

3

+ K

4

A

4

)

Table 8 Multi-wavelength setup options

Options

Description

More

For further Options

Store Off/On

With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.

% Trans/Abs

To switch to % transmittance, concentration or absorbance readings

λ

Wavelength

To enter the measurement wavelength. Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from 320–1100 nm.

Send Data

To send data to a printer, PC or USB memory stick

Concentration Factor

Multiplication factor for converting absorbance values into concentration values.

Concentration
Resolution

To select the position of the decimal point in the calculated concentration readings.

Absorbance Formula

Calculation basis for evaluating samples

Recall Data

Call up saved measurement data, wavelength scans or time courses, see

section 5.3 on

page 28

.

Instrument Setup

Basic data of the instrument, see

section 5.2 on page 17

.